One-Pager

Implanted Cardiac Devices: Pacemakers, ICDs and Wearable Defibrillators

Pacemaker codes and failures, ICD shocks, S-ICD and WCD limits, magnets, cautery and MRI, and discharge teaching

Reviewed by Alex G, BSN, RN, CEN, CCRN, CMC, NREMT-P · Checked against current sources October 2026

Download one-page PDFLetter landscape, ready to print or keep with your notes

Which device does what

DeviceDoesCannot
PacemakerSenses and paces atrium, ventricle or bothShock VT or VF
Leadless pacemakerCapsule in the RV; VVI or VVIRPace the atrium; some ignore a magnet
CRT-P or CRT-DPaces both ventricles (LV lead via coronary sinus); CRT-D also shocksHelp unless it paces nearly every beat
Transvenous ICDATP, cardioversion, defibrillation, backup pacing; stores electrogramsPrevent arrhythmias
S-ICDLead under the skin beside the sternum; shocks VT and VFBradycardia pacing, ATP or CRT (30 s post-shock pacing only)
Wearable defibrillator (WCD)Detects and shocks VT and VF for a limited timePace (see WCD below); protect when not worn

Pacemaker modes

ModeUseWatch for
DDD(R)First choice in sinus node disease; AV block in sinus rhythmMode switch in AF; pacemaker-mediated tachycardia
VVI(R)Slow AF; frail elderly; leadlessPacemaker syndrome (lost AV synchrony)
Lead II · 25 mm/s
Ventricular pacing: every spike is followed by a wide QRS.

Who gets which device

SituationDevice
Symptomatic sinus node dysfunctionPacemaker
Acquired third-degree, high-grade or Mobitz II block, not reversiblePacemaker, even without symptoms
Asymptomatic or sleep-only bradycardiaNo pacemaker; check for sleep apnea (CPAP)
VF arrest or unstable VT, no reversible causeSecondary-prevention ICD whatever the EF, if survival over 1 year expected
Ischemic cardiomyopathy, EF 35% or less, NYHA II to IIIICD at least 40 days after MI and 90 days after revascularization
New cardiomyopathy, or early after MI with very low EFMedical therapy first; a WCD may bridge
HFrEF, EF 35% or less, wide QRS (esp. LBBB), symptomsCRT (CRT-D if an ICD is also indicated)

Pacemaker problems at the bedside

ProblemWhat you seeFirst steps
Failure to captureSpike with no QRS (or no P wave)Check patient and BP; transcutaneous pacing if unstable; device check, chest x-ray
UndersensingSpikes on or right after the patient's own beatsReport (spike-on-T risk); device check
OversensingPauses with no spikes when pacing is expectedRemove the interference; a magnet gives asynchronous pacing in most pacemakers
Failure to paceNo spikes, slow rhythmTreat the bradycardia; urgent device check
Lead II · 25 mm/s
Failure to capture: some spikes are not followed by a QRS.

ICD shocks

SituationAction
One shock, feels fineCall the device clinic for a check that day or the next
Several shocks in a short time, or chest pain, dyspnea, faintingCall 911 or go to the ED
Shocks while awake and talking; sinus tachycardia or noise on the monitorLikely inappropriate: magnet over the ICD, continuous monitoring, external defibrillator ready, call the device team
Electrical storm: 3 or more VA episodes in 24 h, at least 5 min apart, each needing an interventionSedation, nonselective beta-blocker (propranolol) plus amiodarone, fix triggers (ischemia, K, Mg, HF); ablation if it persists

Magnets, cautery and cardioversion

DeviceMagnet on (effect ends when removed)
Most pacemakersAsynchronous pacing (AOO, VOO or DOO) at the maker's magnet rate
Transvenous ICDSuspends tachy detection and shocks in most brands; no asynchronous pacing
S-ICDSuspends shocks; tone for about 60 s
Micra leadlessNo magnet response; reprogram

MRI with a device

Wearable defibrillator (WCD)

After implant: teaching

TopicTeaching
ArmBelow shoulder level for several weeks (4 to 6 on the ICD sheet)
LiftingNothing over 10 to 15 lb, no heavy push or pull, 2 to 3 weeks
IncisionDry 4 to 5 days, then shower; report redness, swelling, drainage, fever
PhonesAt least 6 inches away; opposite ear; not in a chest pocket
Tools, magnetsPower tools and speakers 12 inches away; no magnets near it
IDCarry the card; tell every provider (cautery, MRI, lithotripsy)
Practice Implanted Cardiac Devices: Pacemakers, ICDs and Wearable Defibrillators questions free in your browser26 exam-style questions with rationales. No account or download needed.

In depth

The one-page summary above is what the PDF prints. These sections go further.

Remote monitoring and device data

Pacemaker code, guideline sources and exam points

ModeUseWatch for
AAISinus node disease, intact AV conductionUseless if AV block develops
ModeUseWatch for
VOO or DOOAsynchronous: magnet over a pacemaker, MRI mode or cautery in a dependent patientCompetes with the patient's own beats
SituationUsual device decision
Cardiac sarcoid with a pacing indicationICD with pacing (class IIa in Merck's table)
Young patient, poor venous access or high infection risk, no pacing needS-ICD is an option
LetterMeaningOptions
IChamber pacedA, V, D (both), O (none)
IIChamber sensedA, V, D, O
IIIResponse to a sensed beatI inhibit, T trigger, D both, O none
IVRate responseR raises the rate with activity
VMultisite pacingD (e.g. both ventricles), O none

Unnecessary right ventricular pacing can worsen heart function; programming aims to keep it low (ESC 2021 cites about 20% as an observational threshold).

More bedside detail: pacemaker and ICD problems

ProblemCommon causes
Failure to captureLead dislodgement or fracture, rising threshold, battery depletion
UndersensingLead movement, low-amplitude signals, sensitivity setting
OversensingMuscle noise, EMI, lead fracture noise
Failure to paceOversensing, generator or battery failure, lead disconnection
ProblemWhat you seeCommon causesFirst steps
Twiddler syndromeLoss of capture plus abdominal pulsing or rhythmic arm movementsPatient turns the generator; leads coil and pull backTemporary pacing if needed; x-ray; surgical revision
Pacemaker syndromeDizziness, neck pulsations, dyspnea with VVI pacingLost AV synchronyReport; usually reprogram to a dual-chamber mode

Causes of inappropriate shocks: sinus tachycardia or SVT above the detection rate, AF or flutter, T-wave oversensing (the main S-ICD cause) and noise from a broken lead.

Signs a patient should report at home: dizziness, fainting or near-fainting, shortness of breath, chest pain, return of the old symptoms, or abdominal pulsing or twitching near the device.

More on procedures, MRI, the vest and follow-up

TopicDetail
WCD indication (FDA)At risk of sudden cardiac arrest and not a candidate for, or declining, an implanted defibrillator; e.g. the waiting period after MI, new cardiomyopathy, or an ICD indication when an ICD cannot be placed yet
BystandersStand back and do not touch the patient when the device warns of a shock
Battery and careCharge and change batteries as instructed; answer alerts about the vest
SecurityWalk-through gates are fine without lingering; show the device card and ask for a hand search instead of a wand
Follow-upClinic and remote checks; set up the home transmitter
DrivingRules vary by country and state. NHLBI (US): many states require a 6-month wait after an ICD or WCD placed for an arrest or dangerous arrhythmia, and a high-energy shock can start the same wait; after a device placed to prevent a first arrest a provider may advise about 1 week; no commercial driving with an ICD or WCD. No pacemaker figure was sourced; follow local law and the device team

How the systems are built

A transvenous pacemaker or ICD has a generator in a pocket under the skin or muscle of the upper chest and leads passed through a vein into the right atrium, the right ventricle, or both. CRT adds a third lead that reaches the outside of the left ventricle through the coronary sinus. Modern generators use low-energy circuits and steroid-eluting leads, which helps them last longer. Merck gives 5 to 7 years for ICD generators; MedlinePlus tells patients 6 to 15 years for a pacemaker battery and 4 to 8 years for an ICD. Replacement is a minor operation that swaps the generator and usually keeps the leads.

Leadless pacemakers combine the generator and electrode in a small capsule placed through the femoral vein into the right ventricle and held by tines or a screw. Current devices pace the ventricle only. ESC 2021 notes the lack of randomized comparisons with transvenous single-chamber systems and the difficulty of retrieving a capsule at the end of its life.

The S-ICD places the generator on the side of the chest wall and tunnels a single lead under the skin beside the sternum, so nothing sits inside a vein or the heart. Its shock is larger than a transvenous ICD shock (80 J). Before implant the surface ECG is screened to check that the device will tell the R wave from the T wave; about 8 to 15% of people fail this screen. Lead complications and systemic infection are less common than with transvenous systems, which is why it suits young patients, people with poor venous access, and people at high infection risk such as those on dialysis. It is a poor choice when bradycardia pacing, CRT or antitachycardia pacing for monomorphic VT is likely to be needed.

Modes, timing and pacemaker-specific problems

Merck notes that VVI and DDD give similar survival, but physiologic pacing that keeps the atrium in the loop (AAI, DDD, VDD) may lower the risk of AF and heart failure and slightly improves quality of life. ESC 2021 makes DDD(R) the first choice in sinus node disease and prefers dual-chamber pacing in AV block when the patient is in sinus rhythm, with VVIR reasonable in frail elderly patients or paroxysmal block.

Mode switching protects dual-chamber patients in AF: the device stops tracking the fast atrial rate and paces the ventricle at a programmed rate instead. Pacemaker-mediated tachycardia is a loop in a normal dual-chamber system: a premature or paced ventricular beat conducts back to the atrium, the device senses that atrial signal and paces the ventricle again, and the cycle repeats at a fast rate. The device team reprograms the device to prevent it. Cross-talk is the ventricular channel sensing the atrial pacing pulse and withholding a ventricular beat.

Pacemaker syndrome comes from ventricular pacing without AV synchrony: atria contract against closed valves, causing light-headedness, neck pulsation and dyspnea. The fix is a mode that restores AV timing, most often DDD.

In twiddler syndrome the patient repeatedly rotates the generator in a loose pocket. The leads wind around the generator and their tips pull out of the heart, so capture is lost, and the leads may stimulate the phrenic nerve (diaphragm pacing with abdominal pulsation) or the brachial plexus (rhythmic arm movements). The chest x-ray shows the twisted leads. Older and obese patients with loose tissue are at higher risk; suturing the generator to the fascia helps prevent it. A pacing-dependent patient may need temporary pacing until the leads are repositioned.

Lead impedance is trended at every check, and a sudden change matters more than any single value. Classic teaching: a sudden large rise suggests a conductor fracture or loose connection, while a sudden fall suggests an insulation breach. A fracture also produces bursts of very short, irregular, nonphysiologic signals that an ICD can count as VF, giving shocks while the patient sits in sinus rhythm.

ICD and WCD decisions in more depth

Secondary prevention: after VF or hemodynamically significant VT that was not caused by a reversible or transient problem, the ICD is the preferred treatment (Merck). EF does not decide it. The class III list in Merck's table explains when not to implant: VT or VF from a correctable cause, incessant VT or VF, arrhythmias that ablation can cure, and no reasonable expectation of at least a year of survival with acceptable function.

Primary prevention in ischemic cardiomyopathy waits at least 40 days after MI and 90 days after revascularization, because EF often recovers. The thresholds in Merck's table are EF 35% or less with NYHA class II or III, or EF 30% or less with class I. Nonischemic dilated cardiomyopathy with EF 35% or less and class II or III symptoms qualifies after optimal medical therapy, which the ESC frames as about 3 months. Merck's class IIa list adds cardiac sarcoidosis with EF above 35% plus unexplained syncope, scar on MRI or PET, inducible VT or VF, or a pacing indication.

The wearable defibrillator covers the waiting periods. In the VEST trial (2,302 post-MI patients with EF 35% or less), the primary endpoint of sudden or VT death was not significantly reduced (1.6% vs 2.4%), but all-cause death was lower (3.1% vs 4.9%). ESC 2022 gives the WCD class IIa for adults with a secondary-prevention indication who temporarily cannot have an ICD and class IIb in selected patients early after MI. Shocks are 75 to 150 J, up to five per episode. Observational and registry studies report wear times of 20 to 23.5 hours a day (the VEST median was 18); inappropriate shocks run about 2% in pooled data.

Electrical storm (ESC 2022): ventricular arrhythmia three or more times within 24 hours, separated by at least 5 minutes, each needing an intervention to stop it. Sympathetic drive feeds it, so sedation and a nonselective beta-blocker such as propranolol, usually with amiodarone, come first, along with correcting ischemia, electrolytes and heart failure. Deep sedation and intubation are considered for intractable storm, then autonomic modulation or catheter ablation.

Before any implant, and again as illness advances, the team and patient discuss what the ICD will do near the end of life, including whether and when to turn shocks off.

Procedures, magnets and MRI in more depth

Magnet rates differ by maker (Fernando 2026): 100/min for Abbott and Boston Scientific, 90/min for Biotronik, 85/min for Medtronic and 96/min for Microport, with lower rates when the battery reaches its replacement point. Biotronik's magnet response times out after 8 hours unless the magnet is lifted and replaced. A transvenous ICD does not pace asynchronously with a magnet; it only stops tachy therapy. The S-ICD tone lasts about 60 seconds; if no tone is heard, reprogramming is preferred. The Micra does not respond to a magnet, while the AVEIR switches to asynchronous pacing.

Societies divide surgical risk at the umbilicus, and EHRA/HRS also uses 15 cm from the generator. Bipolar instruments carry essentially no interference risk. With monopolar cautery, lower power, cut rather than coagulation mode, short bursts and a return pad on the opposite lower limb keep current away from the device. The reviewers recommend disabling ICD tachytherapies whenever electrosurgery is used. Afterward the device is interrogated, especially after emergency surgery without a recent check, interference close to the device, a suspected permanent change, or any delivered therapy.

For MRI, Maass and colleagues describe programming a dependent patient to VOO or DOO at 80 to 90/min with high output (RV-only in a biventricular device), and a non-dependent patient with a resting rate above 50/min to VVI or DDI at 30 to 40/min. Rate response, magnet response and ICD therapies are turned off. For non-conditional systems, 1.5 T is preferred. Power-on reset can change the mode and make the device prone to oversensing; a device near battery depletion should not be scanned without a plan.

Implant, anticoagulation and the first weeks

Pocket hematoma occurs in about 2 to 5% of implants, usually within the first week. It matters because it raises the chance of device infection several-fold and can force reoperation and a pause in anticoagulation. In BRUISE CONTROL (patients with an annual clot risk of 5% or more), continuing warfarin gave clinically significant hematoma in 3.5% versus 16.0% with heparin bridging. DOACs are handled by individual risk; dual antiplatelet therapy is a strong bleeding risk factor.

Arm restrictions protect the new leads while scar tissue anchors them. The most common late problems are infection (fever, chills, pocket pain or redness, which can appear months after surgery), lead dislodgement or fracture, and device recalls.

Living with a device: teaching in plain words

Practice free in your browser3,800+ exam-style questions with rationales. No account, no ads. Also free on the App Store and Google Play.

Related one-pagers

Sources

  1. Mitchell LB. Cardiac Pacemakers. Merck Manual Professional Edition, full review February 2025. Link
  2. Mitchell LB. Implantable Cardioverter-Defibrillators (ICD). Merck Manual Professional Edition, full review February 2025. Link
  3. Glikson M, Nielsen JC, Kronborg MB, et al. 2021 ESC Guidelines on cardiac pacing and cardiac resynchronization therapy. Eur Heart J. 2021;42(35):3427-3520. Link
  4. Zeppenfeld K, Tfelt-Hansen J, de Riva M, et al. 2022 ESC Guidelines for the management of patients with ventricular arrhythmias and the prevention of sudden cardiac death. Eur Heart J. 2022;43(40):3997-4126. Link
  5. Fernando RJ, Crimi E, Puttur Rajkumar K, et al. Perioperative management of cardiac implantable electronic devices: a state of the art narrative review. J Thorac Dis. 2026;18(5):536. Link
  6. Maass AH, Hemels MEW, Allaart CP. Magnetic resonance imaging in patients with cardiac implantable electronic devices. Neth Heart J. 2018;26(12):584-590. Link
  7. Duncker D, Garcia R, Bonderman D, et al. A review on wearable cardioverter-defibrillators in clinical practice for sudden cardiac death prevention and heart failure management. Europace. 2025;27(11):euaf257. Link
  8. Guarracini F, Preda A, Bonvicini E, et al. Subcutaneous Implantable Cardioverter Defibrillator: A Contemporary Overview. Life (Basel). 2023;13(8):1652. Link
  9. US Food and Drug Administration. PMA P010030/S056, LifeVest Wearable Cardioverter Defibrillator (ZOLL), approval order and indication, December 17, 2015. Link
  10. Ferrick AM, Raj SR, Deneke T, et al. 2023 HRS/EHRA/APHRS/LAHRS Expert Consensus Statement on Practical Management of the Remote Device Clinic. Europace. 2023;25(5):euad123. Link
  11. Birnie DH, Healey JS, Wells GA, et al. Pacemaker or defibrillator surgery without interruption of anticoagulation (BRUISE CONTROL). N Engl J Med. 2013;368(22):2084-2093 (abstract). Link
  12. Ding S, Pan X. Risk Factors and Preventing Strategies of Pocket Hematoma After Cardiac Implantable Electronic Device Implantation: A Systematic Review. J Cardiovasc Dev Dis. 2025;12(12):490. Link
  13. Mahdi A, Mahdi M, Vasudeva R, et al. Twiddler Syndrome: A Rare Complication of Pacemaker Lead Placement. Kans J Med. 2023;16(2):137-138. Link
  14. Scott J. Device-detected Atrial Fibrillation: How Much is Too Much? Arrhythm Electrophysiol Rev. 2026;15:e20. Link
  15. MedlinePlus. Heart pacemaker - discharge. US National Library of Medicine, reviewed 7/14/2024. Link
  16. MedlinePlus. Implantable cardioverter defibrillator - discharge. US National Library of Medicine, reviewed 7/14/2024. Link
  17. National Heart, Lung, and Blood Institute. Pacemakers: After Getting a Pacemaker and Living With a Pacemaker. Link
  18. National Heart, Lung, and Blood Institute. Defibrillators: Living With an ICD or WCD. Link