Swan-Ganz (PA Catheter): Waveforms, Wedge and Normal Values
Normal pressures and flows, insertion waveforms RA to wedge, safe wedging, end-expiration reads, complications and shock profiles
Download one-page PDFLetter landscape, ready to print or keep with your notes
PA catheter normal values
A Swan-Ganz (pulmonary artery) catheter floats through the right heart into the pulmonary artery to measure right-sided pressures, wedge pressure, cardiac output and mixed venous oxygen saturation. Normal adult values: CVP 2–6, PA 15–30/8–15, wedge (PAOP) 6–12 mm Hg, cardiac index 2.5–4.0 L/min/m², SvO₂ 60–80%.
| Pressure | Normal (mm Hg) | Reflects |
|---|---|---|
| CVP / RA | 2–6 | Right-sided filling |
| RV | 15–30 / 2–8 | RV systolic = PA systolic |
| PA | 15–30 / 8–15 | Pulmonary vascular load |
| PA mean | About 10–22 | Pulmonary hypertension if high |
| Wedge (PAOP, PCWP) | 6–12 | Left atrial and LV filling |
| Flow and resistance | Normal | How it is figured |
|---|---|---|
| Cardiac output | 4–8 L/min | Thermodilution or continuous |
| Cardiac index | 2.5–4.0 L/min/m² | CO ÷ BSA |
| Stroke volume | 60–100 mL | CO ÷ HR × 1000 |
| SVR | 800–1200 dyn·s/cm⁵ | (MAP − CVP) × 80 ÷ CO |
| PVR | Under 250 dyn·s/cm⁵ | (mean PA − PAOP) × 80 ÷ CO |
| SvO₂ | 60–80% | Distal port sample or fiberoptic |
Swan-Ganz waveforms during insertion
| Chamber | Depth (right IJ) | Waveform |
|---|---|---|
| RA | About 20 cm | Low pressure, a and v waves |
| RV | 30–35 cm | Tall systolic, diastolic near 0 and flat or rising |
| PA | 40–45 cm | Diastolic steps up and falls; dicrotic notch after the T wave |
| Wedge | About 50 cm | Low, damped trace with a and v waves |
- The balloon goes up once the catheter is past the sheath (at least 20 cm in) and stays up while advancing; deflate it before pulling back.
- No RV trace by 40–45 cm, or no PA trace by 50–55 cm: deflate, pull back to about 20 cm and try again (likely coiling).
- Ectopy is common as the tip crosses the RV and settles once it reaches the PA. The catheter can cause right bundle branch block, and in a patient with left bundle branch block insertion may cause complete heart block: have pacing ready.
Wedge pressure: technique and safety
- Confirm a clean PA waveform first. If it is damped or already looks wedged, do not inflate.
- Inflate with the catheter syringe only until the trace changes to wedge; never past the 1.5 mL the balloon holds.
- Read at end-expiration on a frozen or printed trace, at the end of the a wave (just before the QRS).
- Deflate as soon as the reading is taken, following your unit's technique, and confirm the PA trace returns. If it does not, the catheter may need pulling back about 2 cm.
- Never flush the distal port while wedged.
Reading at end-expiration
- At end-expiration, pleural pressure is closest to atmospheric, so CVP, PA and wedge are read there in every patient.
- Breathing on their own: just before the pressure dips with inspiration. On the ventilator: just before it rises with the breath.
- The digital number averages the whole breath and can be wrong; use the graphic trace, ideally next to the airway pressure or capnography waveform.
- Level the transducer to the phlebostatic axis and zero it, the same as an arterial line.
- PEEP of 10 cmH₂O or more makes CVP and wedge overread true filling; note the PEEP with every reading.
Cardiac profiles by shock type
| Shock | CI | CVP | Wedge | SVR | SvO₂ |
|---|---|---|---|---|---|
| Hypovolemic | ↓ | ↓ | ↓ | ↑ | ↓ |
| Cardiogenic (LV) | ↓ | ↑ or normal | ↑ | ↑ | ↓ |
| Obstructive (PE, tamponade) | ↓ | ↑ | ↑ in tamponade, normal or low in PE | ↑ | ↓ |
| Distributive (septic, early) | ↑ or normal | ↓ or normal | ↓ or normal | ↓ | ↑ or normal |
PA catheter complications
| When | Complication |
|---|---|
| Insertion | Arrhythmias (usually brief), RBBB or complete heart block, pneumothorax, arterial puncture, air embolism |
| While in place | Line infection (mostly at the sheath), thrombosis, migration |
| Balloon | Pulmonary infarction from a persistent wedge; PA rupture (0.1–0.8%, often fatal); balloon rupture |
PA rupture is more likely with older age, female sex, pulmonary hypertension, anticoagulation, hypothermia (a stiffer catheter) or a catheter advanced too far. Remove the catheter when its numbers no longer change care.
In depth
The one-page summary above is what the PDF prints. These sections go further.
PA catheter ports and lumens
| Port | Sits in | Used for |
|---|---|---|
| Distal | Pulmonary artery | PA and wedge pressure, mixed venous (SvO₂) samples |
| Proximal | Right atrium | CVP, ScvO₂, thermodilution injectate |
| Balloon port | Balloon near the tip | Wedging with the syringe that comes with the catheter; 1.5 mL |
| Thermistor | Near the tip | Blood temperature and thermodilution cardiac output |
Why the wedge reflects the left atrium (and when it does not)
With the balloon up, flow beyond the tip stops and a still column of blood links the catheter to the pulmonary veins and left atrium. That only works where pulmonary arterial and venous pressures exceed alveolar pressure (West zone 3). In zones 1 and 2 the catheter reads alveolar pressure. Low volume and high PEEP enlarge those zones.
- Clues the tip is in the wrong zone: no clear a and v waves, or a wedge that swings widely with each breath.
- Mitral regurgitation gives tall v waves on the wedge; read at the a wave, not the peak.
- Overwedging (balloon against the vessel wall) gives a pressure that keeps rising, is not pulsatile or reads above PA diastolic.
- Catheter whip gives sharp spikes at the start of systole on RV and PA traces; moving the catheter 1–2 cm may help.
Cardiac output and SvO₂ at the bedside
Bolus thermodilution injects cold fluid into the right atrium and the thermistor times its passage; continuous catheters warm the blood with a filament instead. Fast infusions through the sheath or proximal port, tricuspid regurgitation, intracardiac shunts and low injectate volume all disturb the reading. The continuous trend averages several minutes, so it lags behind sudden changes.
SvO₂ shows the balance between oxygen delivery and use. When it falls, check the four parts of delivery and demand: cardiac output, hemoglobin, arterial saturation and oxygen consumption (fever, shivering, pain, work of breathing).
Nursing care of a PA catheter
- Watch the PA waveform continuously; a spontaneous wedge trace or an RV trace means migration.
- Keep the balloon deflated and the syringe attached but empty, or follow your unit's lock-out practice.
- Check the insertion depth at the sheath each shift and after moving the patient; compare with the post-insertion chest film.
- Level and zero with position changes; read at end-expiration; trend numbers with the patient, not alone.
- Sterile sheath and dressing care as for any central line; remove when it no longer guides therapy.
Exam pearls
- Diastolic step-up from RV to PA is how you know the tip crossed the pulmonary valve.
- Wedge is always lower than mean PA. High PA pressures with a normal wedge point to pre-capillary pulmonary hypertension; a high wedge points to the left heart.
- Damped PA trace that looks like a wedge with the balloon down: migration, not a reading.
- Hemoptysis after inflation: stop, call, bleeding side down.
- Cardiogenic shock: low CI, high wedge, high SVR, low SvO₂. Septic shock: high or normal CI, low SVR, high SvO₂.
Related one-pagers
- Hemodynamic NormalsNormal adult values, key formulas and how to read the waveforms behind them
- Shock Types at a GlanceHemodynamic patterns, exam findings and first treatments for each shock type
- Vasopressors and InotropesReceptors, doses, SSC 2026 septic shock order, push-dose pressors and extravasation
- ACS and STEMITerritories, time goals, first drugs and complications
- Heart FailureHFrEF vs HFpEF, stages, GDMT pillars, warm/cold profiles, shock and diuretics
- Right Heart Failure and Pulmonary HypertensionRV physiology, recognition, PH groups, ICU causes, drugs, intubation risk and mechanical support
Calculators for this topic
- Pulmonary Vascular Resistance (PVR)
- Stroke Volume (SV)
- Systemic Vascular Resistance (SVR)
- BSA (Mosteller)
Sources
- Bootsma IT, Boerma EC, de Lange F, Scheeren TWL. The contemporary pulmonary artery catheter. Part 1: placement and waveform analysis. J Clin Monit Comput. 2022;36:5–15. Link
- Bootsma IT, Boerma EC, de Lange F, Scheeren TWL. The contemporary pulmonary artery catheter. Part 2: measurements, limitations, and clinical applications. J Clin Monit Comput. 2022;36:17–31. Link
- Nakamura Y, Saku K, Homma S, et al. Pulmonary artery rupture by pulmonary artery catheter in cardiac surgery: a case report and review of literature. Front Cardiovasc Med. 2025;12:1567723. Link
- Saugel B, Kouz K, Meidert AS, Schulte-Uentrop L, Romagnoli S. How to measure blood pressure using an arterial catheter: a systematic 5-step approach. Crit Care. 2020;24:172. Link